Game Ready Half Leg Boot Use Guide

User guide 4 sayfa · 605.07 KB · PDF

ATX

FOR USE WITH

GAME READY GRPRO $ ^{®} $ 2.1 SYSTEM

MED4 ELITE™ SYSTEM

Image

GAME READY™

Image

个

HALF LEG BOOT

USE GUIDE

ENGLISH

SIZING SPECIFICATIONS

• Size: One Size

• Length: 19"

ASSEMBLY

The Heat Exchanger (PN 520906-03) fits inside the Sleeve (PN 510906) and comes as part of a pre-assembled Wrap (PN 590906-03). If the Heat Exchanger is removed to launder the Sleeve, or for any other reason, please use the following instructions to re-insert the Heat Exchanger into the Sleeve.

Image

心

WARNING

It is mandatory to fully read and understand your System's User Manual before using the device. Failure to follow operating instructions could result in serious injury.

Image

△!

IMPORTANT

1 Place the Sleeve on a flat surface with logo side up, and determine the opening location.

Read complete indications, contraindications, cautions, and warning before using this product. Keep this document for future reference.

Image

ce ne the

3 Be sure the Heat Exchanger is flat inside the Sleeve with no folds or creases. Zip up the Sleeve, if applicable. After assembly, lay the blue side up and ensure the Heat Exchanger is uniformly laid out and smooth inside the Sleeve.

2 Insert the Heat Exchanger into the Sleeve with the blue side down (The blue side of the Heat Exchanger should touch the blue side of the Sleeve).

Image

DAME MFAOY

Image

순간에

REMOVAL OF HEAT EXCHANGER

  1. Disconnect the Connector Hose from the Wrap

  2. Unzip the zipper

  3. Gently pull out the Heat Exchanger

STORAGE OF YOUR WRAP

Hang your Wrap on a wide hanger or lay flat. Do not fold or stack it, as this could kink the fluid chamber and the Wrap will not work properly.

BEFORE INITIAL USE

Prime the Wrap using the following steps:

• With the GRPro 2.1 Control Unit off, attach the Connector Hose to the Control Unit and the Wrap

  • Lay the Wrap open and flat next to the Control Unit (not on the body)

Turn the System on and run for 2 minutes with No Pressure

  • With the Med4 Elite Control Unit off, attach the Connector Hose to Patient 1 on Control Unit and the Wrap

  • Lay the Wrap open and flat next to the Control Unit (not on the body)

Press the On/Off button above the touchscreen

  • Choose Patient 1 and run Cold Therapy with No Pressure for 2 minutes

HALF LEG BOOT WRAP APPLICATION