Autosampler Configuration for Node Control (For Old & New Versions)

Neye ihtiyacın var

  1. Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: adım 1, 3 resimden 1. Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: adım 1, 3 resimden 2. Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: adım 1, 3 resimden 3.
    • 1. Turn on the autosampler. Screen will read "...STANDBY..."

    • 2. Press "ENTER/PROGRAM" button. Use arrows to highlight "CONFIGURE". Press "ENTER/PROGRAM" button

    • This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the "ENTER/PROGRAM" button.

    • 3. Use arrows to find "EVENT MARK". Press the "ENTER/PROGRAM" button.

  2. Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: adım 2, 3 resimden 1. Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: adım 2, 3 resimden 2. Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: adım 2, 3 resimden 3.
    • 4. Use arrow to select "CONTINUOUS SIGNAL". Press "ENTER/PROGRAM" button.

    • 5. Use arrow to select "PUMP CYCLE". Press "ENTER/PROGRAM" button.

    • It is critical this setting is configured in order for the sensor node to operate the autosampler!

    • 6. Use arrows to find "EXIT CONFIGURATION." Press "ENTER/PROGRAM" button. This will bring up the screen "...STANDBY...

Bitiş Çizgisi

Bir başkası bu kılavuzu tamamladı.

Brooke Mason

Üyelik tarihi: 05/29/19

2.928 İtibar

17 adet Kılavuz yazıldı

Ekip

Open-Storm Open-Storm üyesi

Community

14 Üyeler

24 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 3

Son 30 gün: 10

Her zaman: 97