Ana içeriğe geç
  1. 1. Switch battery "ON". 2. Autosampler: Push "ON" button.
    • 1. Switch battery "ON".

    • 2. Autosampler: Push "ON" button.

    • It will say ...STANDBY...

  2. 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM”  button to select "PROGRAM". 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.
    • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

    • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

  3. 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button. 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.
    • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

    • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

    • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

  4. 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED". 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1". 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).
    • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

    • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

    • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

    • It will take some time before it starts sampling.

    • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

Bitiş Çizgisi

Bir başkası bu kılavuzu tamamladı.

Brooke Mason

Üyelik tarihi: 29-05-2019

2.385 İtibar

15 adet Kılavuz yazıldı

Takım

Open-Storm Open-Storm üyesi

Community

15 Üyeler

19 adet Kılavuz yazıldı

0 Comments

Yorum Ekle

İstatistikleri Görüntüle:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 0

Her zaman: 82