  1. 1. Switch battery "ON". 2. Autosampler: Push "ON" button.
    • 1. Switch battery "ON".

    • 2. Autosampler: Push "ON" button.

    • It will say ...STANDBY...

  2. 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM”  button to select "PROGRAM". 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.
    • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

    • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

  3. Android Tamir Setleri

    Yeni bir ekran veya pil, bir tamir kiti kadar yakınında.

    Hemen Alışveriş Yapın

    Android Tamir Setleri

    Yeni bir ekran veya pil, bir tamir kiti kadar yakınında.

    Hemen Alışveriş Yapın
  4. 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button. 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.
    • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

    • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

    • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

  5. 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED". 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1". 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).
    • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

    • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

    • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

    • It will take some time before it starts sampling.

    • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

Bitiş Çizgisi

Bir başkası bu kılavuzu tamamladı.

Brooke Mason

Üyelik tarihi: 29-05-2019

2.289 Reputation

15 adet Kılavuz yazıldı

Team

Open-Storm Open-Storm üyesi

Community

13 Üyeler

19 adet Kılavuz yazıldı

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 2

All Time: 82