Bear 1000 Main Power Receptacle Replacement

Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.

Neye ihtiyacın var

  1. Bear 1000 Main Power Receptacle Replacement, Air and O2 Blender: adım 1, 1 resimden 1.
    • Remove the ribbon cable at the connection on the oxygen blender control board.

  2. Bear 1000 Main Power Receptacle Replacement: adım 2, 1 resimden 1.
    • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.

  3. Bear 1000 Main Power Receptacle Replacement: adım 3, 1 resimden 1.
    • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.

  4. Bear 1000 Main Power Receptacle Replacement, Air and O2 Inlet Manifold: adım 4, 1 resimden 1.
    • Remove the pneumatic and electrical connections labeled.

    • Remove the water trap bowls on the rear of the unit.

  5. Bear 1000 Main Power Receptacle Replacement: adım 6, 1 resimden 1.
    • Remove the four screws and remove the manifold.

    • Remove the two screws holding the power supply to the manifold.

    • Note and mark the orientation of the wires. Remove the two wires going to the fuses at the fuse end and the ground wire going to the main ground stud.

  6. Bear 1000 Main Power Receptacle Replacement: adım 9, 1 resimden 1.
    • Using an allen wrench, remove the mounting screws and nuts from the receptacle.

    • Transfer the three wires form the old receptacle to the new one

Sonuç

To reassemble your device, follow these instructions in reverse order.

Eli Ferman

Üyelik tarihi: 04/08/20

607 İtibar

4 adet Kılavuz yazıldı

Ekip

Cal Poly, Team S7-G12, Paton Fall 2020 Cal Poly, Team S7-G12, Paton Fall 2020 üyesi

CPSU-PATON-F20S7G12

2 Üyeler

6 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 4

Son 30 gün: 6

Her zaman: 102