Neye ihtiyacın var

  1. Fisher Paykel Neopuff Valve System Replacement, Valve System: adım 1, 1 resimden 1.
    • Remove the back cover, fixed by four screws.

    • Disconnect the tube from the manometer.

    • Remove the manometer by unscrewing the two retaining nuts.

    • Fit the existing manometer into the new valve assembly panel.

    • Tighten the retaining nuts.

    • Reconnect the manometer tube.

    • Refit the front panel to the back cover with the four screws from step 1.

  2. Fisher Paykel Neopuff Valve System Replacement: adım 8, 1 resimden 1.
    • Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.

Sonuç

To reassemble your device, follow these instructions in reverse order.