Philips Emerson CoughAssist CA Series Disassembly

Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.

Neye ihtiyacın var

  1. Philips Emerson CoughAssist CA Series Disassembly, Philips Emerson CoughAssist CA Series Disassembly: adım 1, 1 resimden 1.
    • First, remove the two 16mm pan head screws (1) directly under the front panel of the ventilator. (Page 3).

    • Next, remove the two 16mm flat head screws (3) from the back panel of the ventilator. (Page 3).

    • Lift off the top cover (2) of the device. (Page 3).

  2. Philips Emerson CoughAssist CA Series Disassembly: adım 4, 2 resimden 1. Philips Emerson CoughAssist CA Series Disassembly: adım 4, 2 resimden 2.
    • Now remove the "Inhale flow" control knob (4) and "Inhale pressure" knob (4), by loosening the setscrew on each before pulling off. If you have a CM model, carefully remove the Manual Control Lever (3) by prying it off with the screwdriver. (Page 3).

  3. Philips Emerson CoughAssist CA Series Disassembly: adım 5, 1 resimden 1.
    • Unscrew the three 12.7mm plastite screws ( 2) holding the panel assembly to the lower housing. A long Phillips screwdriver is very helpful for access to the three screws. (Page 3)

    • The front control panel (5) should now be loose. Carefully slide it forwards to remove. (Page 4).

Sonuç

To reassemble your device, follow these instructions in reverse order.

Zachary Stednitz

Üyelik tarihi: 04/07/20

292 İtibar

2 adet Kılavuz yazıldı

Ekip

Cal Poly, Team S11-G18, Paton Spring 2020 Cal Poly, Team S11-G18, Paton Spring 2020 üyesi

CPSU-PATON-S20S11G18

3 Üyeler

3 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 0

Son 30 gün: 1

Her zaman: 226