  1. Philips Respironics System One REMstar Blower Unit Replacement, Blower Unit: adım 1, 1 resimden 1.
    • Turn the device upside down so that the rubber feet are facing you.

    • Use the TR15 screw head to remove the two 46.5mm length crews located on the back.

    • Gently remove the casing so that the wires are not damaged.

  2. Philips Respironics System One REMstar Blower Unit Replacement: adım 2, 1 resimden 1.
    • Using a Torx TR10 screwdriver, remove the seven 13.5 mm screws that are securing the inner casing.

    • Remove the inner casing.

  3. Philips Respironics System One REMstar Blower Unit Replacement: adım 3, 1 resimden 1.
    • Remove the motor from the bottom half of the inner casing and set it to the side.

  4. Philips Respironics System One REMstar Blower Unit Replacement: adım 4, 2 resimden 1. Philips Respironics System One REMstar Blower Unit Replacement: adım 4, 2 resimden 2.
    • Remove the circular piece by popping it off to separate the circuit board from the top cover.

    • No tools should be necessary to complete this action.

  5. Philips Respironics System One REMstar Blower Unit Replacement: adım 5, 2 resimden 1. Philips Respironics System One REMstar Blower Unit Replacement: adım 5, 2 resimden 2.
    • Disconnect the green and red wire connectors from their white connectors to disconnect the blower motor.

Sonuç

To reassemble your device, follow these instructions in reverse order.

2 farklı kişi bu kılavuzu tamamladı.

Bir Yorum

These steps may be out of sequential order. Step 4 should probably be Step 1 and perhaps some are even missing.

Charles Marshall - Yanıt

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 1

Son 7 gün: 5

Son 30 gün: 15

Her zaman: 1,214