Philips Respironics V200 Esprit Oxygen Flow Sensor/Valve Replacement

Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.
    • If the ventilator is not already in diagnostic mode, simultaneously press the ALARM RESET and 100% O2 keys on the front panel for approximately five seconds while turning ventilator power on, then touch OK to enter diagnostic mode. (Service Manual 8-19)

    • Reconnect the oxygen source to the ventilator. (Service Manual 8-19)

  1. Philips Respironics V200 Esprit Oxygen Flow Sensor/Valve Replacement: adım 3, 1 resimden 1.
    • Touch Oxygen on the Hardware screen and set the flow to 1 LPM. Verify that the O2Position display reads from 185 to 525 steps. (Service Manual 8-19)

  2. Philips Respironics V200 Esprit Oxygen Flow Sensor/Valve Replacement: adım 4, 1 resimden 1.
    • Touch Oxygen and set the flow to 180 LPM. Check that the O2 Position display corresponds to the set altitude: (Service Manual 8-19)

    • Connect a patient circuit and analyzer to the ventilator. (Service Manual 8-19)

    • Set the analyzer’s function to read LPM mode and gas select to O2 (set a high flow range capable of measuring 180 LPM oxygen). (Service Manual 8-19)

  3. Philips Respironics V200 Esprit Oxygen Flow Sensor/Valve Replacement: adım 7, 1 resimden 1.
    • Check that set oxygen flow corresponds to measured flow: (Service Manual 8-19/20)

Sonuç

To reassemble your device, follow these instructions in reverse order.

Ethan Norman

Üyelik tarihi: 05/04/20

157 İtibar

1 Kılavuz yazıldı

Ekip

Cal Poly, Team S1-G4, Zimmerman Spring 2020 Cal Poly, Team S1-G4, Zimmerman Spring 2020 üyesi

CPSU-ZIMMERMAN-S20S1G4

5 Üyeler

5 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 1

Son 30 gün: 5

Her zaman: 201