Ana içeriğe geç

Eşyalarını Tamir Et

Onarma Hakkı

Parçalar ve Aletler

Biomed
Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.
Biomed

Giriş

This is a replacement guide for the fan in a Philips Respironics V60 ventilator. The cooling fan within the ventilator prevents the system from heating up too much. So it is important to ensure that this fan is kept clean and functioning.

For more information on replacing the fan, please refer to page 9-6 in the service manual.

  1. Press and release the ON/Shutdown key, opening the Shutdown window in the screen. Select Ventilator Shutdown, which will shut down the ventilator.
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Disconnect the internal battery from the power harness connector.
    • Disconnect the internal battery from the power harness connector.

  3. Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Remove the 4 screws using a screwdriver.
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

  5. Remove the fan mount push pins that attach the fan to the back panel.
    • Remove the fan mount push pins that attach the fan to the back panel.

    • Pliers may be required.

  6. Disconnect fan cable from motor controller.
    • Disconnect fan cable from motor controller.

  7. Remove fan,
    • Remove fan,

    • from back panel.

  8. Remove fan filter assembly (the fan filter and housing), from outside the back panel.
    • Remove fan filter assembly (the fan filter and housing), from outside the back panel.

    • When installing, orient the fan with the fan cable at the top inside of the fan, and orient the fan filter and housing as shown.

Sonuç

To reassemble your device, follow these instructions in reverse order.

Rie Kuraoka

Üyelik tarihi: 23-01-2020

814 İtibar

7 adet Kılavuz yazıldı

Takım

Cal Poly, Team S7-G10, Paton Spring 2020 Cal Poly, Team S7-G10, Paton Spring 2020 üyesi

CPSU-PATON-S20S7G10

2 Üyeler

9 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 2

Son 30 gün: 4

Her zaman: 23