Ana içeriğe geç

ResMed VPAP Auto 25 Circuit Board Replacement

Neye ihtiyacın var

  1. ResMed VPAP Auto 25 Circuit Board Replacement, Circuit Board: adım 1, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement, Circuit Board: adım 1, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement, Circuit Board: adım 1, 3 resimden 3.
    • Remove the four 26.7 mm T3 Torx screws from the bottom side of the ResMed VPAP Auto 25.

  2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 2, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 2, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 2, 3 resimden 3.
    • Flip the device over.

    • Pry the outermost plastic cover off with the metal spudger until the cover is loose and able to be removed.

  3. ResMed VPAP Auto 25 Circuit Board Replacement: adım 3, 2 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 3, 2 resimden 2.
    • Pull off the front cover to reveal the internal components.

  4. ResMed VPAP Auto 25 Circuit Board Replacement: adım 4, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 4, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 4, 3 resimden 3.
    • Loosen the motherboard from the rubber dowels holding it in place and unplug the screen components.

  5. ResMed VPAP Auto 25 Circuit Board Replacement: adım 5, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 5, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 5, 3 resimden 3.
    • Remove the three 26.7 mm T3 Torx screws from the now visible rubber fan motor cover.

    • Lift off the rubber fan motor cover, revealing the fan motor.

  6. ResMed VPAP Auto 25 Circuit Board Replacement: adım 6, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 6, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 6, 3 resimden 3.
    • Remove the fan motor unit from its housing.

    • Remove the three 26.7 mm screws securing the motor housing to the bottom piece with a T3 Torx screwdriver.

  7. ResMed VPAP Auto 25 Circuit Board Replacement: adım 7, 3 resimden 1. ResMed VPAP Auto 25 Circuit Board Replacement: adım 7, 3 resimden 2. ResMed VPAP Auto 25 Circuit Board Replacement: adım 7, 3 resimden 3.
    • Pull out the fan motor housing from the bottom cover, revealing a foam insulation pad and the circuit board.

    • Slide the circuit board out of its plastic grooves in the bottom cover to retrieve the desired component.

Sonuç

To reassemble your device, follow these instructions in reverse order.