Ana içeriğe geç

Smart Tweezers and LCR-Reader LCR-meter Tweezer Probes Replacement

Neye ihtiyacın var

  1. Smart Tweezers and LCR-Reader LCR-meter Tweezer Probes Replacement, Removing the old tips: adım 1, 1 resimden 1.
    • Start by removing the old tips with a T8 Torx screw driver.

    • Keep the screws and washers

  2. Smart Tweezers and LCR-Reader LCR-meter Tweezer Probes Replacement, Cleaning and replacing the tips: adım 2, 2 resimden 1. Smart Tweezers and LCR-Reader LCR-meter Tweezer Probes Replacement, Cleaning and replacing the tips: adım 2, 2 resimden 2.
    • After the tips have been removed, it is best to wipe down the arms with isopropyl. This will ensure there are no debris in the way which may cause false readings.

    • Replace the tips using the washers and screws from the old set. If the device shows a reading while not in contact with anything, remove the tips, clean them and the arms again, and replace.

Sonuç

If the device is showing a false reading, remove the tips and clean again.