Ana içeriğe geç

Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement

Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.

Neye ihtiyacın var

  1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement, SPM Vent Kit: adım 1, 1 resimden 1.
    • Loosen but do not remove the(4) 8-32 X 3 screws on the outer air intake.

  2. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 2, 2 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 2, 2 resimden 2.
    • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator.

  3. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 3, 2 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 3, 2 resimden 2.
    • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”.

  4. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 4, 1 resimden 1.
    • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module.

  5. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 5, 3 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 5, 3 resimden 2. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 5, 3 resimden 3.
    • Remove the battery compartment cover by unscrewing the (4) 6-32 X 5/16 screws.

  6. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 6, 1 resimden 1.
    • Remove the battery by unscrewing the (4) 6-32 X 2 ¼ screws and detaching the plug from its lockinglatch.

  7. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 7, 1 resimden 1.
    • Unscrew the (4) 6-32 X 5/16 Phillips screws to remove the damaged battery compartment case.

  8. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 8, 1 resimden 1.
    • Loosen and remove the (2) 4-40 X 1 ¼ Screws and nylon spacers supporting the Power Input assembly unto the chassis

  9. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 9, 1 resimden 1.
    • Loosen and remove the (2) 6-32 X ¼ screws holding the chassis to the dovetail mounting bracket and remove the SPO2 insulator. Loosen and remove the (2) 8-32 X ¼ screws holding the chassis to the vent module

  10. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 10, 1 resimden 1.
    • Insert screwdriver through the holes on the chassis to loosen and remove the (2) 6-32 x 5/16 screws holding the chassis to the vent module.

  11. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 11, 1 resimden 1.
    • Lift the damaged chassis from the ventilator module.

  12. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 12, 1 resimden 1.
    • Disconnect the Power Input cable from the PIM PCB by pressing on the locking latch and pulling the cable straight up from the connector.

  13. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 13, 1 resimden 1.
    • Remove the PIM PCB by loosening the (5) 4-40 X 5/16 screws.

  14. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 14, 1 resimden 1.
    • Remove the Dovetail stabilizer bracket by loosening and removing the (2) 6-32 x ¼ and the (2) 4-40 X ¼ screws.

  15. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 15, 1 resimden 1.
    • Install the dovetail stabilizer bracket ontothe new SPM assembly by tightening the (2) 6-32 X ¼ and (2) 4-40 X ¼ screws.

  16. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 16, 1 resimden 1.
    • Secure the PIM Board to the new SPM with the (5) 4-40 x 5/16 screws.

  17. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 17, 1 resimden 1.
    • Secure the chassis to the vent module using the (2) 6-32 X ¼ screws with SPO2 insulator, the (2) 8-32 X ¼ and the (2) 6-32 x 5/16 screws.

  18. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 18, 1 resimden 1.
    • Secure the Power Input assembly to the Chassis using the (2) spacers and (2) 4-40 X 1 ¼ screws. Do not over-tighten the screws.

  19. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 19, 2 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 19, 2 resimden 2.
    • Rotate battery compartment to mate with upper and lower case cutouts and press firmly into place. Secure with (4) 6-32 X 5/16 screws provided

  20. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 20, 1 resimden 1.
    • Re-assemble the battery by connecting its cable to the connector (pull on cable to insure it is locked in place) then tightening the (4) 6-32 X 2 ¼ screws.

  21. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 21, 3 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 21, 3 resimden 2. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 21, 3 resimden 3.
    • Re-assemble the battery compartment cover by tightening the (4) 6-32 X 5/16 screws.

  22. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 22, 1 resimden 1.
    • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position.

  23. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 23, 1 resimden 1.
    • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10-32 Keps nuts and the (4) 6-32 X 2” screws.

  24. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 24, 2 resimden 1. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 24, 2 resimden 2.
    • Tighten the (4) 8-32 X 3 screws on the outer air intake.

  25. Zoll Impact Uni-Vent 731 SPM Vent Kit Replacement: adım 25, 1 resimden 1.
    • Perform HiPot Testing then perform Calibration and Functional Test using the RCS.

Sonuç

To reassemble your device, follow these instructions in reverse order.

Bir başkası bu kılavuzu tamamladı.

Brandon Drury

Üyelik tarihi: 05/04/20

391 İtibar

1 Kılavuz yazıldı

Ekip

Cal Poly, Team S1-G5, Zimmerman Spring 2020 Cal Poly, Team S1-G5, Zimmerman Spring 2020 üyesi

CPSU-ZIMMERMAN-S20S1G5

5 Üyeler

5 adet Kılavuz yazıldı

Bir Yorum

Hi

What is the service passcode to reset the PM clock?

fayezhmc - Yanıt

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 0

Son 30 gün: 2

Her zaman: 316