The Midmark M11 Steam Sterilizer is designed to be safe, reliable and easy to use. With one of the largest chambers of any tabletop autoclave on the market, the M11 can optimize your supply of available sterile instruments by reducing the time and effort required for sterilization.

1 Soru Tümünü görüntüle

water leaks after cycle

water leaks after cycle completes

Bu soruyu yanıtla Ben de bu sorunu yaşıyorum

Bu iyi bir soru mu?

Puan 0
Yorum Ekle

2 Cevap

Filtre ölçütü:

A small amount of water is normal after a cycle. The door will open to dry, and some steam will condense and drip on the counter. A large amount of water usually indicates a bad door gasket or door dam.

Paul Olson

Rescue Biomedical Services

https://rescuebiomed.com/autoclaves-and-...

Bu yanıt yardımcı oldu mu?

Puan 0
Yorum Ekle

Is the half moon dam gasket in place on the bottom of the door?

Bu yanıt yardımcı oldu mu?

Puan 0
Yorum Ekle

Yanıtını ekle

Nancy Machusick sonsuza kadar minnettar olacak.