Ana içeriğe geç

Repair and service information related to the ResMed AirSense 10 CPAP machine. Also known as the ResMed Air10.

28 Soru Tümünü görüntüle

My humidifier heater isn't working.

My Airsense 10 is about 2 years old. My humidifier heater quit. I pulled the cover and checked the fuse and it is good. I don't have a schematic to check the test points. Could you give me some info on the heater?

Thanks!

Bu soruyu yanıtla Ben de bu sorunu yaşıyorum

Bu iyi bir soru mu?

Puan 0
1 Yorum

Yes this is the fuse under the heater plate. I unclipped it to test its continuity.

tarafından

Yorum Ekle

1 Cevap

Filtre ölçütü:
En Yararlı Yanıt

Bg Higg by checked the fuse I assume your are referring to the one under the heater unit like shown on here Data on the heater is not freely available so I would go ahead and replace the heater on use a bench power supply to see if that could run the heater.

Bu yanıt yardımcı oldu mu?

Puan 1
Yorum Ekle

Yanıtını ekle

Bg Higg sonsuza kadar minnettar olacak.
İstatistikleri Görüntüle:

Son 24 Saat: 1

Son 7 gün: 4

Son 30 gün: 20

Her zaman: 312