  1. Maquet Servo-U O2 Supply Replacement, O2 Supply: adım 1, 1 resimden 1.
    • Turn the inspiratory channel lock one quarter of a turn to release.

    • Lift off the inspiratory channel cover.

    • WARNING! The inspiratory channel cover shall be kept in place once maintenance has been performed. Otherwise the gas modules can be affected and deliver the wrong flow.

    FixBot'a Sor

    Yorum Ekle

  2. Maquet Servo-U O2 Supply Replacement: adım 2, 1 resimden 1.
    • Lower the locking catch.

    FixBot'a Sor

    Yorum Ekle

  3. Maquet Servo-U O2 Supply Replacement: adım 3, 2 resimden 1. Maquet Servo-U O2 Supply Replacement: adım 3, 2 resimden 2.
    • Disconnect the O2 cell connector.

    • Lift and discard the O2 cell with the rubber seal to special waste and/or remove and discard the bacterial filter to hazardous waste.

    FixBot'a Sor

    Yorum Ekle

  4. Maquet Servo-U O2 Supply Replacement: adım 4, 1 resimden 1.
    • Ensure that the new rubber seal is intact, then firmly put in a new bacterial filter.

    FixBot'a Sor

    Yorum Ekle

  5. Maquet Servo-U O2 Supply Replacement: adım 5, 1 resimden 1.
    • Connect the O2 cell connector.

    • Put the O2 cell in position.

    • Close the locking catch.

    FixBot'a Sor

    Yorum Ekle

  6. Maquet Servo-U O2 Supply Replacement: adım 6, 1 resimden 1.
    • Put the inspiratory channel cover in position.

    • Tighten the inspiratory channel lock.

    • Perform a pre-use check stated in Troubleshooting Page.

    FixBot'a Sor

    Yorum Ekle

Bitiş Çizgisi

İptal et: Bu kılavuzu tamamlamadım.

Cassandra Aarrestad

Üyelik tarihi: 01/23/20

627 İtibar

4 adet Kılavuz yazıldı

Ekip

Cal Poly, Team S7-G2, Paton Spring 2020 Cal Poly, Team S7-G2, Paton Spring 2020 üyesi

CPSU-PATON-S20S7G2

3 Üyeler

5 adet Kılavuz yazıldı

0Kılavuz Yorumları

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 7

Son 30 gün: 51

Her zaman: 2,963