ULCO Integrus PSV ventilator O-Ring Replacement

ULCO Integrus PSV ventilator O-Ring Replacement

Preston Myers

Preston Myers ve 3 diğer katkıda bulunanlar

En son May 7, 2024 tarihinde güncellendi

25 dakika
Orta
Devam Ediyor
Tıbbi Ekipman
iFixit de, bu onarım kılavuzu da tıbbi cihaz üreticisi tarafından onaylanmış değildir ya da herhangi birinin üreticiyle alâkası yoktur.

This guide will serve to help access and replace the O-Ring.

Preparation to Service:

  1. Make sure that the anesthetic machine is OFF
  2. Gain access to ventilator housing from this guide here
    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. ULCO Integrus PSV ventilator O-Ring Replacement: adım 4, 1 resimden 1.
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. ULCO Integrus PSV ventilator O-Ring Replacement: adım 5, 1 resimden 1.
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

  3. ULCO Integrus PSV ventilator O-Ring Replacement, O-Ring: adım 6, 1 resimden 1.
    • Pull out the exhaust gas tube (EV910424) and replace the O-ring in the groove on the tube and smear with silicone grease.

  4. ULCO Integrus PSV ventilator O-Ring Replacement: adım 7, 1 resimden 1.
    • Pull out the drive gas tube (EV9104251) and replace the O-ring in the groove on the tube and smear with silicone grease.

  5. ULCO Integrus PSV ventilator O-Ring Replacement: adım 8, 1 resimden 1.
    • Pull out the blow off gas tube (EV910426).

  6. ULCO Integrus PSV ventilator O-Ring Replacement: adım 9, 1 resimden 1.
    • Unscrew the 12 retaining screws on the top plate. Carefully lever the top plate back taking care to ensure that none of the cables connecting it to the body of the ventilator become disconnected.

  7. ULCO Integrus PSV ventilator O-Ring Replacement: adım 10, 1 resimden 1.
    • Using tweezers, remove the O-ring from inside the blow off gas tube receptacle on the flow block. Smear the O-ring groove with silicone grease and insert the new O-ring.

Sonuç

To reassemble your device, follow these instructions in reverse order.

Bir başkası bu kılavuzu tamamladı.

Preston Myers

Üyelik tarihi: 04/09/20

493 İtibar

4 adet Kılavuz yazıldı

Ekip

Cal Poly, Team S7-G13, Paton Spring 2020 Cal Poly, Team S7-G13, Paton Spring 2020 üyesi

CPSU-PATON-S20S7G13

3 Üyeler

8 adet Kılavuz yazıldı

0 Yorum

Yorum Ekle

İstatistikleri Görüntüle:

Son 24 Saat: 0

Son 7 gün: 0

Son 30 gün: 0

Her zaman: 28